310 1220 Infinity II LC System User Manual
11
Maintenance and Repair
Diode Array Detector (DAD)
Maintenance of the Flow Cell
When If the flow cell needs repair due to leaks or contaminations (reduced light throughput)
Tools required Description
Wrench, 1/4 inch
for capillary connections
Hexagonal key, 4 mm
Toothpick
Parts required Description
For parts, see “Standard Flow Cell” on page 352.
Preparations Turn the flow off.
Remove the front cover.
Remove the flow cell, see “Remove and Install a Flow Cell” on page 307.