1220 Infinity II LC System User Manual 205
Error Information
10
Pump Error Messages
index position is reached. If the index position is not recognized within a
defined time, the error message is generated.
Initialization Failed
Error ID: 2207, 2217
The module failed to initialize successfully within the maximum time window.
A maximum time is assigned for the complete pump-initialization cycle. If the
time is exceeded before initialization is complete, the error message is
generated.
Missing Pressure Reading
Error ID: 2054
The pressure readings read by the pump ADC (analog-digital converter) are
missing.
The ADC reads the pressure signal of from the damper every 1ms. If the
readings are missing for longer than 10 s, the error message is generated.
Probable cause Suggested actions
1
Disconnected or defective encoder cable.
Please contact your Agilent service
representative.
2
Defective pump drive assembly.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Blocked (passive or active) inlet valve.
Exchange the (passive or active) inlet valve.
2
Defective pump drive assembly.
Please contact your Agilent service
representative.
3
Defective main board.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1220 Infinity II LC and is the answer not in the manual?

Agilent Technologies 1220 Infinity II LC Specifications

General IconGeneral
BrandAgilent Technologies
Model1220 Infinity II LC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals