General Detector Error Messages
Illegal Temperature Value from Sensor on Main Board
Error ID: 1071
This temperature sensor (located on the detector main board) delivered a
value outside the allowed range. The parameter of this event equals the
measured temperature in 1/100 centigrade. As a result the temperature
control is switched off.
Illegal Temperature Value from Sensor at Air Inlet
Error ID: 1072
This temperature sensor delivered a value outside the allowed range. The
parameter of this event equals the measured temperature in 1/100 centigrade.
As a result the temperature control is switched off.
UV Lamp Current
Error ID: 7450
The UV lamp current is missing.
Probable cause Suggested actions
1
Defective sensor or main board.
Please contact your Agilent service
representative.
2
Detector is exposed to illegal ambient
conditions.
Verify that the ambient conditions are within
the allowed range.
Probable cause Suggested actions
1
The temperature sensor is defect.
• Replace the cable to the main board.
• Please contact your Agilent service
representative.
2
Detector is exposed to illegal ambient
conditions.
Verify that the ambient conditions are within
the allowed range.