294 1220 Infinity II LC System User Manual
11
Maintenance and Repair
Variable Wavelength Detector (VWD)
Replace the Flow Cell / Cuvette Holder
When If application needs a different type of flow cell or the flow cell needs repair.
Tools required Description
Wrench, 1/4 inch
for capillary connections
Parts required # Description
1 Flow cell
Preparations Turn the lamp OFF.
1
Remove the lower front cover.
2
Disconnect the inlet capillary and outlet tubing from the
flow cell.
3
Unscrew the two thumb screws.
4
Pull the flow cell out of its location.