1220 Infinity II LC System User Manual 293
Maintenance and Repair
11
Variable Wavelength Detector (VWD)
9
Replace the fan cover (1.) and fix its screw (2.).
10
Close the lower front cover.
Next Steps:
11
Reset the lamp counter as described in the Utilities software documentation.
12
Turn the l amp ON.
13
Give the lamp more than 10 min to warm-up.
14
Perform Wavelength Calibration to check the correct positioning of the lamp.