1220 Infinity II LC System User Manual 219
Error Information
10
Autosampler Error Messages
Metering Home Failed
Error ID: 4054, 4704
The metering piston has failed to move back to the home position.
The home position sensor on the sampling unit flex board monitors the home
position of the piston. If the piston fails to move to the home position, or if the
sensor fails to recognize the piston position, the error message is generated.
Missing Vial
Error ID: 4019, 4034, 4541, 4706
No vial was found in the position defined in the method or sequence.
When the gripper arm picks a vial out of the sample tray, the processor
monitors the gripper motor encoder. If a vial is present, the closing of the
gripper fingers is limited by the vial. However, if no vial is present, the gripper
fingers close too far. This is sensed by the processor (encoder position),
causing the error message to be generated.
Probable cause Suggested actions
1
Defective sensor or main board.
Please contact your Agilent service
representative.
2
Broken piston.
Exchange the metering piston and seal.
3
Defective metering-drive motor.
Please contact your Agilent service
representative.
4
Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
No vial in the position defined in the
method or sequence.
Install the sample vial in the correct position, or
edit the method or sequence accordingly.
2
Incorrect gripper alignment.
Align gripper.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1220 Infinity II LC and is the answer not in the manual?

Agilent Technologies 1220 Infinity II LC Specifications

General IconGeneral
BrandAgilent Technologies
Model1220 Infinity II LC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals