EasyManuals Logo

Agilent Technologies 1220 Infinity II LC User Manual

Agilent Technologies 1220 Infinity II LC
398 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #50 background imageLoading...
Page #50 background image
50 1220 Infinity II LC System User Manual
3
Installation
Performing a ‘Checkout Run’
Performing a ‘Checkout Run’
1 Start the Chromatographic Data System
2 Create a checkout method with the following parameters:
• Flow: 1 mL/min
• Inj. Volume: 20 μL (Autosampler)
• Oven temp: not controlled
• VWD Wavelength: 254 nm
• Runtime: 1 min
This checkout run is done with the factory installed ‘restriction capillary’ in
place.
3 Prepare 1ml of a checkout sample (acetone for example) and put it on vial
position 1 in the autosampler tray.
For Manual Injector configurations load 20 μL of checkout sample in the
loop. Overfill the injection loop at least 3 times. (e.g. inject at least 60 μL at
a 20 μL sample loop)
4 Start a single run.
As result a single peak should be visible.
5 Print out the report.
6 Store all created and printed out reports in a binder.
The instrument is installed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1220 Infinity II LC and is the answer not in the manual?

Agilent Technologies 1220 Infinity II LC Specifications

General IconGeneral
BrandAgilent Technologies
Model1220 Infinity II LC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals