EasyManuals Logo

Agilent Technologies 1220 Infinity II LC User Manual

Agilent Technologies 1220 Infinity II LC
398 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #207 background imageLoading...
Page #207 background image
1220 Infinity II LC System User Manual 207
Error Information
10
Pump Error Messages
Pressure Above Upper Limit
Error ID: 2014, 2500
The system pressure has exceeded the upper pressure limit.
Pressure Below Lower Limit
Error ID: 2015, 2501
The system pressure has fallen below the lower pressure limit.
5
Defective pump drive assembly.
Please contact your Agilent service
representative.
6
Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
Probable cause Suggested actions
1
Upper pressure limit set too low.
Ensure the upper pressure limit is set to a value
suitable for the analysis.
2
Blockage in the flowpath (after the
damper).
Check for blockage in the flow path. The
following components are particularly subject
to blockage: inline filter frit, needle
(autosampler), seat capillary (autosampler),
sample loop (autosampler), column frits and
capillaries with small internal diameters (e.g.
50 µm ID).
3
Defective damper.
Please contact your Agilent service
representative.
4
Defective main board.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1220 Infinity II LC and is the answer not in the manual?

Agilent Technologies 1220 Infinity II LC Specifications

General IconGeneral
BrandAgilent Technologies
Model1220 Infinity II LC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals