Error Information
VWD Detector Error Messages
Cutoff filter doesn't decrease the light intensity at 250 nm
Error ID: 7813
The automatic filter check after lamp ignition has failed. When the lamp is
switched on, the detector moves the cutoff filter into the light path. If the filter
is functioning correctly, a decrease in lamp intensity is seen. If the expected
intensity decrease is not detected, the error message is generated.
Filter Missing
Error ID: 7816
The filter motor is not detected.
5
Current step value too far from maximum.
• Repeat the calibration.
• Please contact your Agilent service
representative.
6
Misaligned/defective grating assembly.
Please contact your Agilent service
representative.
7
Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
Probable cause Suggested actions
1
Motor is not connected.
Please contact your Agilent service
representative.
2
Defective motor.
Please contact your Agilent service
representative.
3
Defective/missing grating or filter.
Please contact your Agilent service
representative.
4
Cable/connector defective.
Please contact your Agilent service
representative.