222 1220 Infinity II LC System User Manual
10
Error Information
Autosampler Error Messages
Needle Up Failed
Error ID: 4017
The needle arm failed to move successfully from the seat or out of the vial to
the upper position.
The upper position of the needle arm is monitored by a position sensor on the
sampling unit flex board. The sensor detects the successful completion of the
needle movement to the upper position. If the needle fails to reach the end
point, or if the sensor fails to recognize the needle arm movement, the error
message is generated.
Probable cause Suggested actions
1
Needle installed incorrectly, or wrong
needle type (too long).
Ensure the correct needle type is used, and
installed correctly.
2
Defective or dirty position sensor.
Please contact your Agilent service
representative.
3
Defective motor.
Please contact your Agilent service
representative.
4
Sticking spindle assembly.
Please contact your Agilent service
representative.
5
Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Defective or dirty position sensor.
Please contact your Agilent service
representative.
2
Defective motor.
Please contact your Agilent service
representative.