1220 Infinity II LC System User Manual 147
Test Functions and Calibration
9
Solvent Delivery System
Running the Leak Test
Running the test from the Agilent Lab Advisor
1 Select the Leak Test from the Test Selection menu.
2 Start the test and follow the instructions.
When If problems with the pump are suspected
Tools required Description
Wrench, 1/4 inch (supplied in HPLC Tool-Kit)
Parts required # p/n Description
1 G1313-87305 Restriction Capillary
1 01080-83202 Blank nut
1500mL Isopropanol
Preparations Place a bottle of LC-grade isopropyl alcohol in the solvent cabinet and connect its solvent tube
to the active inlet valve of the pump.
NOTE
Make absolutely sure that all parts of the flow path that are part of the test are very
thoroughly flushed with IPA before starting to pressurize the system. Any trace of other
solvents, or the smallest air bubble inside the flow path, will definitely cause the test to
fail.
NOTE
Make sure to release the pressure by slowly opening the purge valve when the test has
finished.
NOTE
Evaluating the Results describes the evaluation and interpretation of the leak test results
(“Evaluating the Leak Test Results” on page 148).
NOTE
For detailed instructions refer to the Agilent Lab Advisor tool.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1220 Infinity II LC and is the answer not in the manual?

Agilent Technologies 1220 Infinity II LC Specifications

General IconGeneral
BrandAgilent Technologies
Model1220 Infinity II LC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals