232 1220 Infinity II LC System User Manual
10
Error Information
VWD Detector Error Messages
Grating Missing
Error ID: 7819
The grating motor is not detected.
No heater current
Error ID: 7453
The lamp heater current in the detector is missing. During lamp ignition, the
processor monitors the heater current. If the current does not rise above the
lower limit within 1, the error message is generated.
Probable cause Suggested actions
1
Grating motor is not connected.
Please contact your Agilent service
representative.
2
Cable/connector defective.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Lamp disconnected.
Ensure the lamp is connected.
2
Ignition started without the top foam in
place.
Please contact your Agilent service
representative.
3
Fan not running (permitting lamp on).
Please contact your Agilent service
representative.
4
Defective main board.
Please contact your Agilent service
representative.
5
Defective or non-Agilent lamp.
Exchange the lamp.
6
Defective power supply.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1220 Infinity II LC and is the answer not in the manual?

Agilent Technologies 1220 Infinity II LC Specifications

General IconGeneral
BrandAgilent Technologies
Model1220 Infinity II LC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals