1220 Infinity II LC System User Manual 57
Installation
3
Install the Restriction Capillary
Install the Restriction Capillary
When
Tools required p/n Description
8710-0510 Wrench open 1/4 — 5/16 inch
Parts required Description
Restriction capillary
Preparations Remove the column, see “Remove a Column” on page 55.
1
Open both front covers.
2
Place the restriction capillary inside the column
compartment.
3
Connect the restriction capillary to the injector valve
(port 6).
4
Connect the restriction capillary to the flow cell inlet of
the detector.