1220 Infinity II LC System User Manual 17
Site Requirements and Specifications
2
Site Requirements
Power Cord
Different power cords are offered as options with the system. The female ends
of all power cords are identical. The female end plugs into the power-input
socket at the rear left side of the instrument. The male end of each power cord
is different and designed to match the wall socket of a particular country or
region.
Bench Space
The dimensions and weight of the Agilent 1220 Infinity II LC allow it to be
placed on almost any desk or laboratory bench. It needs an additional 2.5 cm
(1.0 in) of space on either side and approximately 8 cm (3.1 in) at the rear for
air circulation and electric connections.
Make sure that the bench intended to carry the Agilent 1220 Infinity II LC is
designed to bear the weight of the instrument.
The Agilent 1220 Infinity II LC should be operated upright.
Absence of ground connection or use of unspecified power cord
The absence of ground connection or the use of unspecified power cord can lead to
electric shock or short circuit.
âž” Never operate your instrument from a power outlet that has no ground connection.
âž” Never use a power cord other than the Agilent Technologies power cord designed
for your region.
Use of cables not supplied by Agilent
Using cables that have not been supplied by Agilent Technologies can lead to
damage of the electronic components or personal injury.
âž” Never use cables other than the ones supplied by Agilent Technologies to ensure
proper functionality and compliance with safety or EMC regulations.