EasyManuals Logo

Agilent Technologies 1220 Infinity II LC User Manual

Agilent Technologies 1220 Infinity II LC
398 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #226 background imageLoading...
Page #226 background image
226 1220 Infinity II LC System User Manual
10
Error Information
General Detector Error Messages
General Detector Error Messages
These errors are specific to both detector types VWD and DAD.
Heater at fan assembly failed
Error ID: 1073
Every time the deuterium lamp or the tungsten lamp (DAD only) is switched
on or off a heater self-test is performed. If the test fails an error event is
created. As a result the temperature control is switched off.
Heater Power At Limit
Error ID: 1074
The available power of the heater reached either the upper or lower limit. This
event is sent only once per run. The parameter determines which limit has
been hit:
0 means upper power limit hit (excessive ambient temperature drop).
1 means lower power limit hit (excessive ambient temperature increase).
Probable cause Suggested actions
1
Defective connector or cable.
Please contact your Agilent service
representative.
2
Defective heater.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Excessive ambient temperature change.
Wait until temperature control equilibrates.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1220 Infinity II LC and is the answer not in the manual?

Agilent Technologies 1220 Infinity II LC Specifications

General IconGeneral
BrandAgilent Technologies
Model1220 Infinity II LC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals