210 1220 Infinity II LC System User Manual
10
Error Information
Pump Error Messages
The sensor on the main board detects when the top foam is in place. If the
module is restarted with the foam removed, the module switches off within
30 s, and the error message is generated.
Servo Restart Failed
Error ID: 2201, 2211
The pump motor in the module was unable to move into the correct position
for restarting.
When the module is switched on, the first step is to switch on the C phase of
the variable reluctance motor. The rotor should move to one of the C positions.
The C position is required for the servo to be able to take control of the phase
sequencing with the commutator. If the rotor is unable to move, or if the C
position cannot be reached, the error message is generated.
Probable cause Suggested actions
1
The module started with the top cover and
foam removed.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Mechanical blockage of the module.
Remove the pump-head assembly. Ensure there
is no mechanical blockage of the pump-head
assembly or pump drive assembly.
2
Disconnected or defective cable.
Please contact your Agilent service
representative.
3
Blocked (passive or active) inlet valve.
Exchange the (passive or active) inlet valve.
4
Defective pump drive assembly.
Please contact your Agilent service
representative.
5
Defective main board.
Please contact your Agilent service
representative.