200 1220 Infinity II LC System User Manual
10
Error Information
General Error Messages
Leak Sensor Short
Error ID: 0082
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak sensor current to
change within defined limits. If the current increases above the upper limit,
the error message is generated.
Remote Timeout
Error ID: 0070
A not-ready condition is still present on the remote input. When an analysis is
started, the system expects all not-ready conditions (for example, a not-ready
condition during detector balance) to switch to run conditions within one
minute of starting the analysis. If a not-ready condition is still present on the
remote line after one minute the error message is generated.
Probable cause Suggested actions
1
Defective leak sensor.
Please contact your Agilent service
representative.
2
Leak sensor incorrectly routed, being
pinched by a metal component.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Not-ready condition in one of the
instruments connected to the remote line.
Ensure the instrument showing the not-ready
condition is installed correctly, and is set up
correctly for analysis.
2
Defective remote cable.
Exchange the remote cable.
3
Defective components in the instrument
showing the not-ready condition.
Check the instrument for defects (refer to the
instrument’s documentation).