1220 Infinity II LC System User Manual 239
Error Information
10
DAD Detector Error Messages
Wavelength Recalibration Lost
Error ID: 1037
The calibration information needed for your detector to operate correctly has
been lost.
During calibration of the detector the calibration values are stored in ROM. If
no data is available in the spectrometer ROM, the error message is generated.
DSP Not Running
This error message comes up when the communication between the optical
unit and the main board has a problem.
No Run Data Available In Device
In a very rare case the capacity of the CompactFlash Card is not sufficient.
This could happen for example when the interrupt of LAN communication
Probable cause Suggested actions
1
The detector is new.
Recalibrate the detector.
2
The detector has been repaired.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Random communication error.
• Switch the detector off and on again at the
power switch. If the error reoccurs:
• Please contact your Agilent service
representative.
2
Defective detector main board.
Please contact your Agilent service
representative.
3
Defective PDA/optical unit.
Please contact your Agilent service
representative.