1220 Infinity II LC System User Manual 291
Maintenance and Repair
11
Variable Wavelength Detector (VWD)
Exchanging the Deuterium Lamp
When If noise or drift exceeds application limits or lamp does not ignite.
Tools required Description
Screwdriver, Pozidriv #1 PT3
Parts required # p/n Description
1 G1314-60100 Deuterium lamp
Preparations Turn the lamp OFF.
WARNING
Injury by touching hot lamp
If the detector has been in use, the lamp may be hot.
âž” If so, wait for lamp to cool down.
1
Remove the lower front cover.
2
Locate the heater fan cover.