EasyManua.ls Logo

Agilent Technologies 1220 Infinity II LC

Agilent Technologies 1220 Infinity II LC
398 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
300 1220 Infinity II LC System User Manual
11
Maintenance and Repair
Variable Wavelength Detector (VWD)
3
Insert the cuvette with the sample into the holder. The
clear side of the cuvette must be visible.
4
Reset the lamp counter as described in the user
interface documentation.
5
Turn the lamp ON.
6
Give the lamp more than 10 min to warm-up.
7
Perform Wavelength Verification/Calibration to check
the correct positioning of the lamp.
8
Replace the bracket and fix the cuvette.
Next Steps:
9
Install the cuvette holder in the instrument.
10
Perform your verification.
/LJKWSDWK
&OHDUVLGH

Table of Contents

Related product manuals