76 1220 Infinity II LC System User Manual
4
LAN Configuration
Automatic Configuration with BootP
5 Make these entries for the instrument:
• MAC address
• Host name, Enter a Hostname of your choice.
The Host Name must begin with "alpha" characters (i.e. LC1260)
• IP address
• Comment (optional)
• Subnet mask
• Gateway address (optional)
The configuration information entered is saved in the Tab File.
6 Click OK.
7 Leave Edit BootP Addresses by pressing Close.
8 Exit BootP Settings by pressing OK.
9 After each modification of the BootP settings (i.e. EditBootPSettings) a stop
or start of the BootP service is required for the BootP service to accept
changes. See “Stopping the Agilent BootP Service” on page 77 or
“Restarting the Agilent BootP Service” on page 78.
10 Power cycle the Instrument.
OR
If you changed the IP address, power cycle the instrument for the changes
to take effect.
11 Use the PING utility to verify connectivity by opening a command window
and typing:
Ping 192.168.254.11 for example.
The Tab File is located at
C:\Documents and Settings\All Users\Application Data\Agilent\BootP\TabFile