Oven
6850
6850
Released: April 2004 6850 Series Gas Chromatograph Service Procedures 237
WARNING
All fittings and tubing should be stainless
steel. The pressure inside a liquid CO
2
tank
can be as high as 1000 psi.
Liquid CO
2
is available in high-pressure tanks
containing 50 pounds of liquid. The tank must
have an internal dip tube or eductor tube to
deliver liquid CO
2
instead of gas. Do not use a
padded tank (one to which another gas is added
to increase the delivery pressure).
1. Connect a CGA 320 to 1/4-inch male NPT
fitting to the tank followed by a Swagelok
part no. SS-200-7-4 1/4-inch female NPT to
1/8-inch tube fitting.
2. Connect the tank to the cryogenic filter inlet
with 1/8-inch diameter heavy-wall stainless
steel tubing. The tubing may be up to 50 feet
long.
Allow some slack in the line to make move-
ment of the tank or GC easier. Coil and
fasten the ends of the tubing to keep it from
“whipping” if it breaks.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series and is the answer not in the manual?

Agilent Technologies 6850 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals