Troubleshooting
6850
6850
Released: April 2004 6850 Series Gas Chromatograph Service Procedures 335
Extraneous peaks
Air peaks are sometimes seen in a chromatogram
when leakage occurs because the valve rotor
does not seal properly. These leaks may not be
detectable using the soap-bubble method.
If a leak is suspected but cannot be located with
soap bubbles, a pressure check will determine
definitely if a leak exists. Extraneous peaks can
occur due to contamination or improper condi-
tioning of the valve. If leaks are not apparent,
clean or condition the valve.
Other causes, totally unrelated to the valve, may
produce similar symptoms. Impure carrier gas
(that is, containing water) can cause extraneous
peaks.
Peak broadening and tailing
Voids in the flow system (valve and connecting
tubing) cause tailing and peak broadening. Use
inlets and liners with small internal diameters
and connect the valve to the inlet or column with
short lengths of connecting tubing of narrow
inner diameter.
If early-eluting peaks are too broad, stationary
phase or thermal focusing effects should be used
with packed-column ports or increased split
flows when capillary split inlets are used. Inlets
should be equipped with narrow inner diameter
liners, and narrow-bore connecting tubing should
be used between the valve and inlet.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series and is the answer not in the manual?

Agilent Technologies 6850 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals