Released: April 2004 6850 Series Gas Chromatograph Service Procedures 158
Replacement procedures
Thermal conductivity detector
TCD cell
1. Open the lid.
2. Remove the column.
3. If installed, remove the TCD capillary
adapter using a 7/16-inch wrench.
4. Open the lid to the service position. See
“Raising the lid to the service position” on
page 1-17.
Capillary adapter
Heater/Sensor wires
PRT wires
Filament wires
TCD filament block