Inlets and valves
6850
6850
Released: April 2004 6850 Series Gas Chromatograph Service Procedures 71
13. Reinstallation is the reverse of removal.
When done, check for leaks using the Control
Module.
Direct valve-to-column connector kit
Use this kit to connect your 6850 gas or liquid
sampling valve directly onto a column to avoid
unacceptable peak tailing. To use this kit, your
6850 GC must have a purged packed inlet.
This kit contains:
Description Part no. Qty.
Direct-tube assembly n/a 1
SS bulkhead union, 1/16-inch to
530 µm column
0100-1515 1
SS bulkhead union, 1/16-inch to
320 µm column
0100-1527 1
Ferrule, Vespel/graphite, 1/16-inch 0100-1512 4
Liner, polyamide, for 530 µm
columns
0100-1513 1
Liner, polyamide, for 320 µm
columns
0100-1514 1
SS nut, 1/16-inch 0100-1511 2
1/8-inch to 1/16-inch reducing
union
0100-1542 1
Liner tool 18900-20850 1