Released: April 2004 6850 Series Gas Chromatograph Service Procedures 16
Replacement procedures
Back cover
WARNING
Shock hazard. Before beginning, turn off the
instrument and disconnect the power cord.
1. Disconnect all cables from the back of the
instrument.
2. Turn off all supply gases at their sources,
then disconnect all gases from the bulkhead
connectors on the back of the instrument.
3. Remove the seven T-20 Torx screws on the
back cover.
4. Replace the cover. Then, reinsert and tighten
the screws.