Inlets and valves
6850
6850
Released: April 2004 6850 Series Gas Chromatograph Service Procedures 115
Cryo fault
Shuts down the inlet temperature if it does not
reach setpoint in 16 minutes of continuous cryo
operation. Note that this is the time to reach the
setpoint, not the time to stabilize and become
ready at the setpoint.
Shutdown behavior
Both Cryo time-out and Cryo fault can cause cryo
shutdown. If this happens, the inlet heater is
turned off and the cryo valve closes. The GC
beeps and displays this message:
Front Inlet Cryo Shutdown
The inlet heater is monitored to avoid overheat-
ing. If the heater remains on at full power for
more than 2 minutes, the heater is shut down.
The GC beeps and displays this message:
Front inlet heating too
slowly; temperature shut off.
To recover from either condition, turn the GC off,
then on, or enter a new setpoint.
Leak testing
Preparation
1. Cool the column to ambient, and cool inlets
to below 75°C to avoid damage to deacti-
vated glass liners.
2. Remove the column from the inlet fitting on
the inside of the oven.
3. If a septum head is installed, and the quality
of the septum (or microseal) and Graphpack-
3D ferrule on the glass liner are unknown,
replace them now.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series and is the answer not in the manual?

Agilent Technologies 6850 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals