EasyManuals Logo

Agilent Technologies 6850 Series User Manual

Agilent Technologies 6850 Series
337 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #260 background imageLoading...
Page #260 background image
G2629A Control Module
6850
6850
Released: April 2004 6850 Series Gas Chromatograph Service Procedures 261
A typical SERVICE method
Before beginning work on an instrument, use the
Control Module to make sure that the SERVICE
method accomplishes the following:
Any heated zone (inlet, detector, oven, valve
box) where you will be working is set to
< 40°C or OFF
All hydrogen gas flows are turned OFF for
any inlet, detector, or valve you will service
or disconnect
All gas flows that will be disconnected are
turned OFF
Any other feature that could be hazardous or
waste resources (such as cryo cooling) is
turned off
Update firmware
To update GC, injector, or Control Module firm-
ware, first disconnect the Control Module from
the GC. Then, install the PC Card containing the
update into the Control Module and reconnect
the Control Module to the GC. Do not update
firmware during a run.
1. Select Status / Service / Update.
2. Select GC Update, Injector Update, or Mod
Update.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series and is the answer not in the manual?

Agilent Technologies 6850 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals