4-8 | IPS-4 Dual Bench (UV / IR) Analyzer
Cell (heated Cell is optional)
Displays the real-time temperature (°C) of the Measuring Cell, as mea-
sured by the Cell RTD (mounted on the Cell, inside the Oven/Sample
System Enclosure). This parameter is available only for applications
with a heated Oven.
Manifold Pressure
Displays the real-time pressure (psia, or other unit) measured at the
Manifold Block. The manifold pressure is used for valve actuation and
for zero flush (if applicable).
Aspirator Pressure
Displays the real-time pressure (psia, or other unit) being used to
drive the Aspirator system (gas applications only).
Cell Pressure
Displays the real-time pressure (psia, or other unit) being measured in
the Measuring Cell.
Tec Temperature
Displays the temperature of the NDIR sensor.
Probe Temperature
Displays temperature at the sample probe.
Sample/Vent line Temperature
Displays temperature of sample/vent bundle.