Diagnostics Screen
The Diagnostics screen (Figure 4-25) allows the user to disable normal
operations while troubleshooting or setting up external control systems.
Figure 4-25.
Diagnostics Screen.
Diagnostics
Current Output Diagnostics
Relay Diagnostics
Valve Diagnostics
Lamp Diagnostics
UV Diagnostic
IR Diagnostic
HOME ALARMS STATUS SETUP
Figure 4-27.
Current Output
Diagnostics Screen.
Current Output 1 Diagnostics
Diagnostics State O󰀨
Current Output 1 61%
Type 4-20 mA
Over Range
Disabled
HOME ALARMS STATUS SETUP
Figure 4-26.
Current Output
Diagnostics Screen.
Current Output Diagnostics
> Current Output 1 Diagnostics
Current Output 2 Diagnostics
Current Output 3 Diagnostics
Current Output 4 Diagnostics
HOME ALARMS STATUS SETUP
Current Output Diagnostics
Test analog outputs.
Relay Diagnostics
Force relay contacts on or off to test the hardware.
Figure 4-28.
Relay Diagnostics
Screen.
Relay Diagnostics
Diagnostics State O󰀨
Relay 1 On
Relay 2 On
Relay 3 On
Relay 4 On
HOME ALARMS STATUS SETUP

Table of Contents

Other manuals for Ametek IPS-4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek IPS-4 and is the answer not in the manual?

Ametek IPS-4 Specifications

General IconGeneral
BrandAmetek
ModelIPS-4
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals