Specications | 2-5
Environmental
Ambient Temperature
-20 °C to +50 °C (-4 °F to +122 °F), without external heating or cooling.
* When liquids are used, ambient temperature is greater than the freezing
point.
Humidity
0–90 % Relative Humidity
Pollution Degree
Pollution Degree 2
Maximum Altitude
2000 metres
Installation Category
Installation Category II
Enclosure Material
Stainless Steel main enclosures. Aluminum disconnect enclosure.
Ingress Protection
IP65 (Zone 1) and NEMA Type 4 (Div 1)
Physical Dimensions
Height: 660 mm (26"), analyzer only
Width: 947 mm (37.3"), analyzer only
Depth: 254 mm (10"), analyzer only
Weight: 86 kg (188 lb), without Backpan
(may vary, depending on system)
Approximately 129 kg (285 lb), with Backpan