Controller / User Interface | 4-19
Figure 4-15.
Flush Times Settings
screen.
Flush Times
Zero 10 sec
Span 30 sec
Sample 20 sec
HOME ALARMS STATUS SETUP
Flush Times
The Flush Times screen (Figure 4-15) displays all of the available flush
times.
Zero flush time
Defines the duration (seconds) the analyzer will flush the sample sys-
tem with Zero fluid during the analyzer’s normal operating sequence.
Span flush time
Defines the duration (seconds) the analyzer will flush the sample sys-
tem with Span fluid during the analyzer’s normal operating sequence.
Sample flush time
Defines the duration (seconds) the analyzer will flush the sample sys-
tem with sample fluid before it will begin measuring.