EasyManuals Logo

Ametek IPS-4 User Manual

Ametek IPS-4
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #199 background imageLoading...
Page #199 background image
Modbus Communication Interface | 8-1
MODBUS COMMUNICATION INTERFACE
This chapter describes the customer serial communication interface on
the IPS-4 Analyzer. The communication protocol implemented is Modicon
Modbus as defined in “Modicon Modbus Protocol Reference Guide” (PI-
MBUS-300 RevJ). The Modbus protocol transmission mode implemented
is Remote Terminal Unit (RTU) with the analyzer operating as a slave
device.
The IPS-4 Analyzer supports RS-485 serial communication standards. The
physical communication connection between an IPS-4 analyzer and a cus-
tomer DCS\SCADA\PLC\DAS or a general-purpose computer is RS-485.
The analyzer RS-485 connection supports 2-wire multi-drop systems.
Hardware:
The jumpers on the CPU board J9 connector should be installed between
the 1-2 and 3-4 pins.
The converter terminal marked TD(B) which is same as Tx+ should be
connected to the CCB_J14 connector marked Tx+ (pin 3) on the Custom-
er Connection Board, and TD(A) should be connected to Tx- (pin 4).

Table of Contents

Other manuals for Ametek IPS-4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek IPS-4 and is the answer not in the manual?

Ametek IPS-4 Specifications

General IconGeneral
BrandAmetek
ModelIPS-4
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals