5-6 | IPS-4 Dual Bench (UV / IR) Analyzer
 
Every Six Months Sheltered Systems
If your analyzer system is installed in a custom shelter, check the
air lters for the shelter purge and air conditioning systems and
replace them if necessary. Depending on the location of the site
and its environmental conditions, it may be necessary to check and
replace these lters more frequently.
Every Six Months Sample/Vent Lines
Inspect the Sample Line and Vent Line for sags, sharp bends, or
damage to the outer skin. If necessary, take appropriate safety pre-
cautions and replace the lines. Perform a leak check on all associ-
ated ttings after replacing any line(s).
Every Year Measuring Cell
Clean the Measuring Cell and the rest of the analyzer sample sys-
tem. This schedule is a minimum requirement. If other conditions
are found to be present, as outlined in this chapter under “Measur-
ing Cell Preventive Maintenance,” cleaning may be required more
frequently.
At the same time, replace the o-rings (always) and windows and/or
window/mirror combinations (if chipped, cracked, or scratched).
3–5 Years Xenon Flash Lamp Assembly
Replace the Xenon Flash Lamp Assembly. The assembly may need
to be replaced sooner if information from the analyzer diagnostics
indicates related alarms. Refer to “Replacing the Xenon (UV) Flash
Lamp Assembly” in this chapter.
3–5 Years Infrared Source Assembly
Replace the Infrared Source Assembly. The assembly may need to
be replaced sooner if information from the analyzer diagnostics indi-
cates related alarms. Refer to “Replacing the Infrared (IR) Source
Assembly” in this chapter.
Expo Technologies MiniPurge
®
System With eTimer (Optional)
Preventive Maintenance Schedule
Frequency Task
If your analyzer uses an Expo Technologies MiniPurge
®
System
With eTimer, refer to the system’s manual for maintenance frequen-
cy and procedures. This manual is shipped with the analyzer.
To test the pressure of the Electronics Enclosure, use the Cabinet
Pressure Port on the Solenoid Block.
3 Years 
®
System (if analyzer is equipped)
Replace the Battery Pack and perform the commissioning tests as de-
scribed in the Expo MiniPurge
®
Type X / ET Size 1 Manual (ML 422).
See “Expo Technologies MiniPurge
®
System With eTimer Spare
Parts” in Chapter 6 for spare parts ordering information.

Table of Contents

Other manuals for Ametek IPS-4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek IPS-4 and is the answer not in the manual?

Ametek IPS-4 Specifications

General IconGeneral
BrandAmetek
ModelIPS-4
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals