2-4 | IPS-4 Dual Bench (UV / IR) Analyzer
Electrical Requirements
Power Consumption
Without Oven Heater
< 500 W maximum start-up (from a cold start) with continuous average,
depending on ambient temperature.
With Oven Heater
< 900 W maximum start-up (from a cold start) with continuous average,
depending on ambient temperature.
Heated Sample Line
Depending on voltage and length (see customer wiring diagram). Max
rating 240 Vac 20 A.
Supply Voltage
120 VAC (105–132 VAC), 47–63 Hz
OR
240 VAC (209–264 VAC), 47–63 Hz
Cell Construction
Gas and Liquid Cells available, in a variety of materials (body can be 316
stainless steel, Hastelloy C-276, or Monel with Sapphire or UV-grade
Quartz windows), to suit a variety of applications.
Pressure Input Signals
Two pressure input signals.
Manifold Block: 0–100 psig
Measuring Cell: (0–30 psia, 0–100 psia, or 0–500 psia)
On-Board Temperature Sensor
The electronics temperature is measured by a solid-state temperature sen-
sor located on the MCU board.