EasyManua.ls Logo

Ametek IPS-4

Ametek IPS-4
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-26 | IPS-4 Dual Bench (UV / IR) Analyzer
Relay Setup Screens
The Relay Setup screens (Figure 4-19a and 4-19b) allow the user to view or
set up the parameters for up to eight relays. For example, Concentration
Alarms can be set up from this screen (Relay Function option).
Figure 4-19a.
Relay Setup screen.
Relays
Relay 1
Relay 2
Relay 3
Relay 4
Relay 5
HOME ALARMS STATUS SETUP
Figure 4-19b.
Relay Setup screen
(Relay 1).
Relay 1
Function Disable
Source None
Value 0
Delay 0 Sec N
Normal Oper Normally Open
HOME ALARMS STATUS SETUP
Relay 1–8 Function
The analyzer can use up to eight relays to indicate the operational sta-
tus of the analyzer. Each relay provides one SPST (Form A) dry (poten-
tial free) contact. The relays are energized (closed) on start-up. Relay
Function options include:
Disable
Select Disable to turn off the functionality of a relay.
Calibrating
The relay will be de-energized if the analyzer is in a calibration
state (Span Flush, Span, Zero Flush, or Zero).
Data Valid
The relay will be de-energized when the analyzer is in any state
other than the Measure state.
Fault
The relay will be de-energized if any system fault alarm condition
is triggered within the analyzer diagnostic system. The analyzer
requires service.
The relay will reset automatically upon correction of the fault
alarm.

Table of Contents

Other manuals for Ametek IPS-4

Related product manuals