Cautions
Review and follow these Cautions to avoid damaging the equipment.
The electronic circuit boards and other static-sensitive components should be stored and
transported in static-shielding carriers or packages.
For electrical-shock protection, the analyzer must be operated from a grounded power source
that has a securely connected protective-ground contact.
If it becomes necessary to handle any of the electronic circuit boards, do not subject the
boards to static discharge. The ideal solution is a static-safe work area. Since such areas
typically are not available at analyzer installation sites, the use of a wrist strap connected
directly to a ground is recommended. If a wrist strap is not available, you should at the very
least touch the metal chassis (to ground yourself) before handling or touching the boards.
When handling the Xenon Flash Lamp, it is very important not to touch the lamp windows
because residual oils from the ngers will absorb ultraviolet light. The window is the at sur-
face at the end of the narrow glass tube. The lamp assembly is fragile and should be handled
with care.
!
CAUTION
!
CAUTION
!
CAUTION
!
CAUTION
xii | IPS-4 Dual Bench (UV / IR) Analyzer