EasyManuals Logo

Ametek IPS-4 User Manual

Ametek IPS-4
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #92 background imageLoading...
Page #92 background image
3-44 | IPS-4 Dual Bench (UV / IR) Analyzer
3. If the sample response time is adequate (first reading after a Zero
function and after the analyzer System State switches from “Zero
to “Measure” is less than 30 seconds to T90), no further adjustment is
necessary.
The sample response time may vary, depending on Sample Line
length.
4. Record these signals in a log book every month.

The Parameter List (Factory Settings) file is shipped with each analyzer.
These configuration sheets list all of the settings applicable to the factory
analyzer (to meet customer requirements). The files are provided on a CD
and located in the “Supplemental Information” section of this manual and
in the Documentation Package shipped with the analyzer.
If any changes are made to the original configuration, be sure to record
the changes for later reference. If the MCU board is ever replaced, this
data must be re-entered in order to override the factory defaults.
If necessary, refer to the Parameter List (Factory Settings) to revert the
analyzer back to its original settings.
NOTE

Table of Contents

Other manuals for Ametek IPS-4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek IPS-4 and is the answer not in the manual?

Ametek IPS-4 Specifications

General IconGeneral
BrandAmetek
ModelIPS-4
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals