4-38 | IPS-4 Dual Bench (UV / IR) Analyzer
The Backup/Restore screen (Figure 4-33) allows you to back up the cur-
rent analyzer configuration settings, restore previously saved analyzer
configuration settings, or restore analyzer configuration parameters to
their original factory-shipped settings.
If major changes are made to the analyzer configuration settings, it is
important to make a backup copy of these settings. This file can then
be used to restore all parameters to their most recently used settings
or the factory-default settings in the event of a serious internal system
error.
Figure 4-33.
Settings screen.
►Backup settings
Restore settings
Restore factory settings
HOME ALARMS STATUS SETUP
NOTE
Backup settings
Saves the current analyzer configuration settings to a file that can later
be used to restore these settings. The settings are stored in a file called
“user.xml.” See “Backing Up Analyzer Configuration Settings” for
details about how to save configuration settings to a file.
Restore settings
Retrieves and restores the most recently saved analyzer configuration
settings from the “user.xml” file. See “Restoring Analyzer Configura-
tion Settings” for details about how to restore configuration settings.
Restore factory settings
Retrieves and restores the original factory-set analyzer configuration
settings from the “factory.xml” file. See “Restoring Analyzer Configu-