4-18 | IPS-4 Dual Bench (UV / IR) Analyzer
Timer interval
Defines the time between each Auto-Calibration cycle. The unit is
determined by the Unit setting.
Unit
Use this to set the time units that will be used to determine the fre-
quency of each Auto-Calibration function. The unit can be set to hours
or days. This function works in conjunction with the Timer interval.
Start Time
Use this to set the start Auto-Calibration in military time.
To set up calibration functions:
1. Select the parameter you wish to change and press ENTER.
2. Use the UP/DOWN arrow keys to view the options available for each
parameter.
3. Change the timer durations (Auto cal timer interval, Zero flush time,
Span flush time):
• Use the UP/DOWN arrow key to increase/decrease the duration
in increments of one (1.00). When using this method, the value
counts up/down from the current setting. This works best for mi-
nor changes.
• If a value exists, use the LEFT arrow key as a backspace to remove
(erase) the characters to the left of the cursor ( ‘_’ ) if applicable,
then use the numeric keys to enter the new value. This works best
for drastic changes.