4-28 | IPS-4 Dual Bench (UV / IR) Analyzer
To change the settings for a relay:
1. Enter a password to allow changes to be made to the relay settings.
2. Select one of the Relay (1–8) sub-menus. Press ENTER to view the pa-
rameters for that relay.
3. Select the required parameter (Relay Function, Source, Value, or Delay),
change the settings, and press ENTER.
Repeat this step for each parameter and each relay.
When entering a value for the Relay Value and Delay parameters:
• Use the UP/DOWN arrow key to increase/decrease the value in
increments of one (1.00). This works best for minor changes. The
relay value is displayed on the right side of the screen.
• If a value exists, use the LEFT arrow key as a backspace to erase the
characters to the left of the cursor ( ‘_’ ) if applicable, then use the
numeric keys to enter the new value. This works best for drastic
changes.