4-30 | IPS-4 Dual Bench (UV / IR) Analyzer
Type
This allows you to select the output value range, either 4–20 mA DC
or 0–20 mA DC for the output. Select ‘0’ for 4–20 mA DC or ‘1’ for 0–20
mA DC.
Over Range
When enabled, output readings of 3.8 and 20.5 mA will occur if the
analyzer exceeds 4-20 mA.
To change the settings for a current output:
1. Enter a password to allow changes to be made to the current output
settings.
2. Select one of the Current Output (1–8) sub-menus. Press ENTER to
view the Current Output screen.
3. Select the required parameter ( Source, Full, Zero, Type, or Over Range),
change the settings, and press ENTER.
Repeat this step for each parameter and each output.
When entering a value for the Full and Zero parameters:
• Use the UP/DOWN arrow key to increase/decrease the value in
increments of one (1.00). This works best for minor changes. The
range is 0–100. The current zero-scale set point is displayed on the
right side of the screen.
• If a value exists, use the LEFT arrow key as a backspace to erase the
characters to the left of the cursor ( ‘_’ ) if applicable, then use the
numeric keys to enter the new value. This works best for drastic
changes.