5-30 | IPS-4 Dual Bench (UV / IR) Analyzer
Condition/Alarm Name Description and Suggested Corrective Action
This alarm condition occurs if the Relay board cannot communicate with
the MCU board due to incompatible rmware versions.
Corrective Action:
The Relay board and/or MCU board rmware will have to be replaced:
• Contact AMETEK for assistance.
This alarm condition occurs if the Display board cannot communicate
with the MCU board due to incompatible rmware versions.
Corrective Action:
The Display board and/or MCU board rmware will have to be replaced:
• Contact AMETEK for assistance.
File system error This alarm condition occurs if the Detector data le (also known as the
DSP, or digital signal processing, le) is missing or has become corrupt.
Corrective Action:
• Contact AMETEK for assistance.
Calibration Required This warning is generated after a span calibration failure or a zero
calibration failure.
Corrective Action:
Perform a successful span calibration:
• Contact AMETEK for assistance.
Zero Cal Required This warning is generated after a zero calibration failure or if a zero has
not been performed.
Corrective Action:
Perform a successful zero calibration:
• Contact AMETEK for assistance.
Zero Cal Failure This warning is generated after a zero calibration failure.
Corrective Action:
Perform a successful zero calibration:
• Contact AMETEK for assistance.
Span Cal Failure This warning is generated after a span calibration failure.
Corrective Action:
Perform a successful span calibration:
• Contact AMETEK for assistance.