Bias Tee, Option 10 10-6 Bias Tee Menus (Field Measurements View)
Vector Network Analyzer MG PN: 10580-00289 Rev. K 10-11
Bias Tee Setup Menu
Bias Tee Menu
Bias Tee (On/Off): Press this soft key to open the “Bias Tee Menu” to turn
On or Off the Bias Tee voltage.
Int voltage P2: Press this soft key to set the internal bias tee voltage that is
directed onto the center conductor of Port 2. The available range is from
12.0 V to 32.0 V in increments of 0.1 . Use the arrow keys, the rotary knob, or
the number keypad to change the setting. When using the number keypad,
press the soft key for voltage units (V), or press the Enter key. Press the Esc
key to exit without changing the setting.
Int Current Limit P2: Press this soft key to set the internal bias tee current
limit for the voltage that is set at Port 2. The available range is from 0 mA to
450 mA in steps of 1 mA. This current limit sets the trip point for the bias tee
for this port. When using the number keypad, the soft key menu displays
2 choices for units: A or mA. Use the arrow keys, the rotary knob, or the
number keypad to change the setting. When using the number keypad, press
the soft key for current units (A or mA), or press the Enter key to use mA.
Press the Esc key to exit without changing the setting.
Back: Press this soft key to the Sweep menu.
Figure 10-10.Bias Tee Setup Menu – MS20xxB and S412E
Off: Press this soft key to turn Off the Bias Tee function.
On: Press this soft key to turn ON the Bias Tee function.
Back: Press this soft key to return to the “Bias Tee Setup Menu” without
changing the current Bias Tee setting.
Figure 10-11. Bias Tee Menu – MS20xxB and S412E
Back
Bias Tee
Off
Int Current Limit P2
## mA
Int Voltage P2
##.# V
Bias Tee Setup