C-3 Self Test or Application Self Test Error Messages Error Messages
C-2 PN: 10580-00289 Rev. K Vector Network Analyzer MG
C-3 Self Test or Application Self Test Error Messages
Self Test
To run self test, press Shift and System (8) and then Self Test. Refer to the results window in
Figure C-1, which summarizes the status of several key functions in the instrument that are
common to all applications (note that your instrument display may differ from this image). If
any subtest shows FAILED, then check that the battery level is adequate for operation, or
check that the temperature is within acceptable limits. Reset to factory defaults with either
Factory Defaults (Esc+On), or Master Reset (System+On).
Caution
Use of Master Reset (System+On), will erase all user saved setups and
measurement traces and will return the instrument to a full Factory Default
condition. If the error persists, then contact your Anritsu Service Center.