6-3 Domain Setup Menus VNA View Menus
6-16 PN: 10580-00289 Rev. K Vector Network Analyzer MG
Gate Setup Menu (continued)
Center Gate: Press this soft key to set the center Gate in either time or
distance units, depending on whether the active trace is in the Time Domain
or the Distance Domain. In time domain, the units are in seconds (s),
milliseconds (ms), microseconds (µs), nanoseconds (ns), or
picoseconds (ps). In distance domain, the units are in meters (m),
centimeters (cm), or millimeters (mm). If the “Application Options Menu (VNA
Mode)” menu is used to set units in feet rather than meters, then the distance
settings are in feet (ft) only.
Span Gate: Press this soft key to set the span Gate in either time or distance
units, depending on whether the active trace is in the Time Domain or the
Distance Domain. In time domain, the units are in seconds (s),
milliseconds (ms), microseconds (µs), nanoseconds (ns), or
picoseconds (ps). In distance domain, the units are in meters (m),
centimeters (cm), or millimeters (mm). If the “Application Options Menu (VNA
Mode)” menu is used to set units in feet rather than meters, then the distance
settings are in feet (ft) only.
Gate Notch
On Off: Press this soft key to turn On or Off the Gate Notch feature. When
Gate Notch is set to OFF, the polarity of the Gate is set to keep everything
between start and stop. When the Gate Notch is set to ON, the polarity of the
Gate is set to reject everything between start and stop.
Gate Shape: Press this soft key to open the “Gate Shape Menu”. The Gate
Shape value is displayed on the key face, and may be Minimum, Nominal,
Wide, or Maximum.
Back: Press this soft key to return to the previous menu.
Figure 6-15. Gate Setup Menu (continued)
Center Gate
# ns
Stop Gate
# ns
Start Gate
# ns
Span Gate
# ns
Gate Function
Off Display On
Gate
Gate Notch
On Off
Gate Shape
Minimum
Back