EasyManua.ls Logo

Anritsu MS2024B - Calibration Menu

Anritsu MS2024B
256 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VNA View Menus 6-4 Calibration Menus
Vector Network Analyzer MG PN: 10580-00289 Rev. K 6-19
Calibration Menu
Key Sequence: Shift, Calibrate (2)
Start Cal: Press this soft key to open the “Next Step” List Box, which contains
the instructions: Connect cal component, select step, and press Enter to
measure:” See Figure 6-24 on page 6-25 for examples. Only for MS20xxC
instruments, this key changes to “Thru Update or Start Cal” when a completed
calibration includes a Thru measurement. See the soft key image at the
bottom of this list.
Cal Type: Press this soft key to open the selection list box, then select a
calibration type. A soft key offers a toggled selection of Flex or Standard.
Refer to “Cal Type Menu”. Also refer to Section “Calibration Types”
on page 6-37. See Figure 6-19 on page 6-21 for Cal Type examples.
Cal Method: Press this soft key to open the selection list box, then select a
calibration method. See Figure 6-22 on page 6-24 for examples. This soft key
is displayed only for MS20xxC instruments..
Cal Line Setup: Press this soft key to open the “Calibration Line Setup
Menu” on page 6-27. A soft key (Cal Line Type) in the “Calibration Line Setup
Menu” on page 6-27 offers a toggled selection of Coax or Waveguide. This
soft key is displayed only for MS20xxC instruments.
DUT Port Setup: Press this soft key to open the “DUT Port Setup Menu
(Coax)” on page 6-29 or the “DUT Port Setup Menu (Waveguide)”
on page 6-33 if the Cal Line Type is set to Waveguide..
Interpolation
On Off: Press this soft key to toggle interpolation On or Off. In Cal Type
Flex mode, Interpolation is always On. After performing a Standard
calibration, you can turn Interpolation On and then change the frequency
range (smaller) or reduce the number of data points. Refer to “Interpolation”
on page 6-22.
Existing Cal Info: Press this soft key to open the “Existing Calibration
Information” List Box. Press Enter or Esc to close the list box.
Cal Correction
On Off: Press this soft key to toggle calibration correction On and Off.
Thru Update or Start Cal: This soft key is displayed in place of the Start Cal
soft key when a completed calibration includes a Thru measurement.
Pressing this key opens the “Calibration Menu with Thru” on page 6-25, which
offers a Thru update or a new calibration. This soft key is displayed only for
MS20xxC instruments.
Figure 6-18. Calibration Menu
DUT Port Setup
Existing Cal Info
Start Cal
Cal Line Setup
Cal Method
SOLT
Cal Type
Full 2 Port
Cal Correction
On Off
Calibration
Interpolation
On Off
Thru Update
or
Start Cal
Cal Type
Flex Standard

Table of Contents

Related product manuals