127
3.3.1.15. Occupied Bandwidth
3.3.1.15.1. MT8820C
[Measurements]
1. [PCC/SCC] Execute OBW_AVG 20 to set the average count for Occupied Bandwidth to 20 times.
2. [PCC/SCC] Execute TESTPRM TX_MAXPWR_Q_F to set Test Parameter to TX1 – Max. Power (QPSK/FullRB).
3. [PCC/SCC] Execute SWP to measure the Occupied Bandwidth (OBW).
4. [PCC/SCC] Execute OBW? to read the OBW measurement result.
5. [PCC/SCC] Execute OBWPASS? to check that the OBW Pass/Fail judgment is Pass.
3.3.1.15.2. MT8821C
[Measurements]
1. Execute OBW_AVG 20 to set the average count for Occupied Bandwidth to 20 times.
2. Execute TESTPRM TX_MAXPWR_Q_F to set Test Parameter to TX1 – Max. Power (QPSK/FullRB).
3. Execute SWP to measure Occupied Bandwidth (OBW).
4. Execute OBW? PCC to read the PCC OBW measurement result.
5. Execute OBWPASS? PCC to check that the PCC OBW Pass/Fail judgment is Pass.
6. Execute OBW? SCC1 to read the SCC–1 OBW measurement result.
7. Execute OBWPASS? SCC1 to check that the SCC–1 OBW Pass/Fail judgment is Pass.