EasyManua.ls Logo

Anritsu MT8820C - Page 126

Anritsu MT8820C
268 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
126
3.3.1.14. In-band Emissions for non-allocated RB
3.3.1.14.1. MT8820C
[Acceptable Value Setting]
1. [PCC/SCC] Execute MOD_AVG 20 to set the average count of Modulation Analysis to 20 times.
2. [PCC/SCC] Execute TP_INBANDE_GEN_D -57.0 to set General Pass/Fail judgment of TX1 IBE/LEAK @
0dBm.
[Measurements]
3. [PCC/SCC] Execute TESTPRM TX_0DBM to set Test Parameter to TX1 IBE/LEAK @ 0 dBm.
4. [PCC/SCC] Set UL RMC Number of RB and Starting RB.
5. [PCC/SCC] Execute SWP to measure the Modulation Analysis.
6. [PCC/SCC] Execute INBANDE_GEN? MAX to read the InBand Emissions (General) measurement result.
7. [PCC/SCC] Execute INBANDE_IMG? MAX to read the InBand Emissions (IQ Image) measurement result.
8. [PCC/SCC] Execute INBANDE_LEAK? MAX to read the InBand Emissions (Carrier Leakage) measurement
result.
9. [PCC/SCC] Execute INBANDEPASS? to check that the InBand Emissions Pass/Fail judgment is Pass.
10. [PCC/SCC] Execute TESTPRM TX_M30DBM to set Test Parameter to TX1 IBE/LEAK @ 30 dBm.
11. Execute steps 4 to 9.
12. [PCC/SCC] Execute TESTPRM TX_M40DBM_Q_P to set Test Parameter to TX1 EVM/IBE/LEAK @ 40 dBm
(QPSK/PartialRB).
13. Execute steps 4 to 9.
3.3.1.14.2. MT8821C
[Acceptable Value Setting]
1. Execute MOD_AVG 20 to set the average count of Modulation Analysis to 20 times.
2. Execute TP_INBANDE_GEN_D -57.0 to set General Pass/Fail judgment of TX1 IBE/LEAK @ 0dBm.
[Measurements]
3. Execute TESTPRM TX_0DBM to set Test Parameter to TX1 IBE/LEAK @ 0 dBm.
4. Set PCC and SCC1 UL RMC Number of RB and Starting RB.
5. Execute SWP to measure the Modulation Analysis.
6. Execute INBANDE_GEN? MAX,PCC to read the PCC InBand Emissions (General) measurement result.
7. Execute INBANDE_IMG? MAX,PCC to read the PCC InBand Emissions (IQ Image) measurement result.
8. Execute INBANDE_LEAK? MAX,PCC to read the PCC InBand Emissions (Carrier Leakage) measurement
result.
9. Execute INBANDEPASS? PCC to check that the PCC InBand Emissions Pass/Fail judgment is Pass.
10. Execute INBANDE_GEN? MAX,SCC1 to read the SCC -1InBand Emissions (General) measurement result.
11. Execute INBANDE_IMG? MAX,SCC1 to read the SCC-1 InBand Emissions (IQ Image) measurement result.
12. Execute INBANDE_LEAK? MAX,SCC1 to read the SCC-1 InBand Emissions (Carrier Leak) measurement
result.
13. Execute INBANDEPASS? SCC1 to check that the SCC1 InBand Emissions Pass/Fail judgment is Pass.
14. Execute TESTPRM TX_M30DBM to set Test Parameter to TX1 IBE/LEAK @ 30 dBm.
15. Execute steps 4 to 13.
16. Execute TESTPRM TX_M40DBM_Q_P to set Test Parameter to TX1 EVM/IBE/LEAK @ 40 dBm
(QPSK/PartialRB).
17. Execute steps 4 to 13.

Table of Contents