125
3.3.1.13. Carrier Leakage
3.3.1.13.1. MT8820C
[Measurements]
1. [PCC/SCC] Execute MOD_AVG 20 to set the average count of Modulation Analysis to 20 times.
2. [PCC/SCC] Execute TESTPRM TX_0DBM to set Test Parameter to TX1 IBE/LEAK @ 0 dBm.
3. [PCC/SCC] Set UL RMC Number of RB and Starting RB.
4. [PCC/SCC] Execute SWP to measure the Modulation Analysis.
5. [PCC/SCC] Execute CARRLEAK? MAX to read the Carrier Leakage measurement result.
6. [PCC/SCC] Execute CARRLEAKPASS? to check that the Carrier Leakage Pass/Fail judgment is Pass.
7. [PCC/SCC] Execute TESTPRM TX_M30DBM to set Test Parameter to TX1 IBE/LEAK @ 30 dBm.
8. Execute steps 3 to 6.
9. [PCC/SCC] Execute TESTPRM TX_M40DBM_Q_P to set Test Parameter to TX1 EVM/IBE/LEAK @ 40 dBm
(QPSK/PartialRB).
10. Execute steps 3 to 6.
3.3.1.13.2. MT8821C
[Measurements]
1. Execute MOD_AVG 20 to set the average count of Modulation Analysis to 20 times.
2. Execute TESTPRM TX_0DBM to set Test Parameter to TX1 IBE/LEAK @ 0 dBm.
3. Set PCC and SCC1 UL RMC Number of RB and Starting RB.
4. Execute SWP to measure the Modulation Analysis.
5. Execute CARRLEAK? MAX,PCC to read the PCC Carrier Leakage measurement result.
6. Execute CARRLEAKPASS? PCC to check that the PCC Carrier Leakage Pass/Fail judgment is Pass.
7. Execute CARRLEAK? MAX,SCC1 to read the SCC1 Carrier Leakage measurement result.
8. Execute CARRLEAKPASS? SCC1 to check that the SCC1 Carrier Leakage Pass/Fail judgment is Pass.
9. Execute TESTPRM TX_M30DBM to set Test Parameter to TX1 IBE/LEAK @ 30 dBm.
10. Execute steps 3 to 8.
11. Execute TESTPRM TX_M40DBM_Q_P to set Test Parameter to TX1 EVM/IBE/LEAK @ 40 dBm
(QPSK/PartialRB).
12. Execute steps 3 to 8.